Intermittent hypoxemia during hemodialysis: AI-based identification of arterial oxygen saturation saw-tooth pattern
BMC Nephrology volume 26, Article number: 214 (2025)
Abstract
Background
Maintenance hemodialysis patients experience high morbidity and mortality, primarily from cardiovascular and infectious diseases. It was discovered recently that low arterial oxygen saturation (SaO2) is associated with a pro-inflammatory phenotype and poor patient outcomes. Sleep apnea is highly prevalent in maintenance hemodialysis patients and may contribute to intradialytic hypoxemia. In sleep apnea, normal respiration patterns are disrupted by episodes of apnea because of either disturbed respiratory control (i.e., central sleep apnea) or upper airway obstruction (i.e., obstructive sleep apnea). Intermittent SaO2 saw-tooth patterns are a hallmark of sleep apnea. Continuous intradialytic measurements of SaO2 provide an opportunity to follow the temporal evolution of SaO2 during hemodialysis. Using artificial intelligence, we aimed to automatically identify patients with repetitive episodes of intermittent SaO2 saw-tooth patterns.
Methods
The analysis utilized intradialytic SaO2 measurements by the Crit-Line device (Fresenius Medical Care, Waltham, MA). In patients with an arterio-venous fistula as vascular access, this FDA approved device records 150 SaO2 measurements per second in the extracorporeal blood circuit of the hemodialysis system. The average SaO2 of a 10-second segment is computed and streamed to the cloud. Periods comprising thirty 10-second segments (i.e., 300 s or five minutes) were independently adjudicated by two researchers for the presence or absence of SaO2 saw-tooth pattern. We built one-dimensional convolutional neural networks (1D-CNN), a state-of-the-art deep learning method, for SaO2 pattern classification and randomly assigned SaO2 time series segments to either a training (80%) or a test (20%) set.
Results
We analyzed 4,075 consecutive 5-minute segments from 89 hemodialysis treatments in 22 hemodialysis patients. While 891 (21.9%) segments showed saw-tooth pattern, 3,184 (78.1%) did not. In the test data set, the rate of correct SaO2 pattern classification was 96% with an area under the receiver operating curve of 0.995 (95% CI: 0.993 to 0.998).
Conclusion
Our 1D-CNN algorithm accurately classifies SaO2 saw-tooth pattern. The SaO2 pattern classification can be performed in real time during an ongoing hemodialysis treatment, provide timely alert in the event of respiratory instability or sleep apnea, and trigger further diagnostic and therapeutic interventions.
Background
Maintenance hemodialysis (HD) patients suffer from unacceptable high morbidity and mortality, with cardiovascular disease (CVD) being the main cause of hospitalization and death. The association between low arterial oxygen saturation (SaO2) during dialysis and poor patient outcomes is well established [1, 2]. Physiologically, the transport of oxygen to peripheral tissues and organs depends on cardiac output, hemoglobin concentration, and saturation of hemoglobin molecules with oxygen. In HD patients, heart failure is a well-recognized cause of tissue hypoxia. Less appreciated is peripheral hypoxia due to structural and functional pathologies of the respiratory system, resulting in reduced SaO2. Next to cardiac output and hemoglobin concentration, SaO2 is a crucial determinant of the amount of oxygen delivered to tissues and organs. Clinically, SaO2 can be assessed using both invasive and non-invasive methods. Recent advancements have introduced technologies that enable continuous monitoring of SaO2 in the HD extracorporeal blood circuit during a treatment session. These innovations have proven valuable for identifying intradialytic hypoxemia [2], monitoring COVID-19 patients [3], and tracking other important clinical outcomes [1] such as respiratory instability [4]. Nocturnal hypoxemia is associated with cardiovascular complications in incident hemodialysis patients [5]. Prolonged intradialytic hypoxemia (PIH), defined as SaO2 below 90% for more than one third of the HD treatment time, is associated with increased hospitalization and mortality rates. In addition, maintenance HD patients with PIH show a pro-inflammatory phenotype, as indicated by higher inflammatory markers, lower serum albumin levels, and resistance to erythropoietin. Furthermore, PIH is more prevalent in HD patients with comorbidities such as congestive heart failure (CHF) and chronic obstructive pulmonary disease (COPD) [1]. Intradialytic hypoxemia is also associated with peridialytic hypertension, possibly related to sympathetic activation and endothelin-1 secretion [6].
Recently, we observed distinct intermittent oscillatory saw-tooth pattern (STP) of SaO2 in patients on HD [4]. In some instances, through detailed clinical observation, we were able to link STP to sleep-related breathing disturbances. This finding is significant, as sleep-disordered breathing is common in dialysis patients [7]. A study found that 37% of males and 34% of females with end-stage kidney disease (ESKD) treated with dialysis had sleep apnea syndrome (SAS). SAS was significantly associated with age, obesity, diabetes, hypertension, and other sleep disorders [8, 9]. These findings highlight the importance of monitoring SaO2 during HD, as early identification of and intervention for intradialytic hypoxemia could mitigate its impact on long-term morbidity and mortality in this vulnerable population.
A recent study using recurrence analysis quantifies an optimal recurrence threshold (εopt) to identify STP [4]. Constructing εopt time series with a rolling window allows real-time detection of intermittent high-frequency high-amplitude STP. εopt correlates highly with SaO2 desaturation density, and, using oxygen desaturation index as a surrogate for the apnea–hypopnea index, εopt accurately distinguishes SAS events.
One-dimensional convolutional neural networks (1D-CNN) have gained significant attention in the medical field, particularly for analyzing time series of biomedical signals such as electrocardiograms (ECG) and electroencephalograms (EEG). These networks can efficiently process time-series data and other one-dimensional formats without extensive pre-processing. Applications include patient-specific ECG classification and early diagnosis of diseases [10].
Quasi-continuous measurement of SaO2 in the extra-corporeal circuit by the Crit-Line monitor (CLM) affords the opportunity to follow the temporal evolution of SaO2 during HD. Against this backdrop, we aimed to automatically identify repetitive episodes of intermittent STP using 1D-CNNs.
Methods
In HD patients with an arterio-venous fistula as vascular access, the oxygen saturation in the extracorporeal HD circuit resembles SaO2. SaO2 during HD was measured using the CLM (Fresenius Medic Care North America, Waltham, MA), a device certified by the US Food and Drug Administration (FDA) for the measurement of hematocrit and oxygen saturation in the extracorporeal blood circuit of the HD system. CLM measures oxygen saturation and hematocrit 150 times per second. These measurements are then averaged over a 10-second period and the respective means transmitted to Amazon Web Services (AWS) via Apache KAFKA, a real time streaming software [11].
Patient selection
To build our AI model, we collected intradialytic SaO2 recordings from HD patients. The inclusion criteria were (a) being an in-center maintenance hemodialysis patient, and (b) availability of CLM SaO2 recordings. The patients were dialyzed in clinics of the Fresenius Kidney Care network that comprises about 2,500 clinics in the US. In about 600 of these clinics, CLM is used as a standard of care. There were no exclusion criteria. These rather wide inclusion criteria and the absence of exclusion criteria ensure that the base population is typical for a US HD population and that the results are generalizable. In these about 600 clinics, we have collected data from about 500,000 HD treatments. Out of these treatments we selected 89 treatments from 22 patients.
Visual inspection
We segmented the treatment time series into 5-minute consecutive sections, each segment had 30 SaO2 measurements. The 5-minute segments were visually inspected and independently adjudicated by two researchers who categorized them as either no saw-tooth pattern (NSTP, stable and consistent SaO2 with minimal fluctuations; Fig. 1A) or STP. NSTP indicates respiratory stability. In contrast, STP shows much greater variability and fluctuations with noticeable peaks and troughs over time; it indicates respiratory instability (Fig. 1B). The visual classification of SaO₂ patterns by the two independent observers was based on two characteristics: amplitude and duration. NSTP were considered SaO₂ variations with an amplitude of less than 2%, with only occasional isolated spikes exceeding 5% in amplitude and lasting less than 30 s. In contrast, STP were considered by oscillations in SaO₂ with an amplitude greater than 5% and a duration exceeding 30 s. Severe STP had an SaO₂ < 90%. The reason for the < 90% SaO₂ criterion was that earlier work showed that intradialytic SaO₂ < 90% is associated with increased morbidity and mortality [1]. Discrepancies between the two observers were resolved through a joint review.
Training
We randomly allocated the 5-minute SaO2 time series segments to either training (80%) or test (20%) data sets. Initially, most segments in the training data were NSTP. To address this class imbalance in the training data, we up-sampled the STP group to match the number of NSTP samples. This was done by randomly selecting STP segments with replacement from the original pool and replacing the original STP samples in the training dataset with the newly selected ones. The test data set was not changed. The updated training dataset was then shuffled and randomly partitioned into two subsets. To improve model generalization, random Gaussian noise (mean = 0, standard deviation = 0.1) was added to every SaO₂ measurement in one subset, which was then used to train the 1D-CNN. The other subset served as the validation set for computing the validation error throughout the optimization process.
For the learning process we used the Keras open-source deep learning library [12] to build a 1D-CNN. The input layer of our 1D-CNN receives the 5-minute time-series segments of SaO2. The first convolutional layer applies 32 filters, each with a kernel size of 6, to extract low-level temporal patterns from the input sequences. A Batch Normalization layer follows the convolution to normalize activations and accelerate training. A Max Pooling layer with a pool size of 3 and a stride of 2 is applied to reduce spatial dimensions while retaining the most important features. A second convolutional layer is applied to the pooled output to refine feature extraction. This layer has 32 filters with a kernel size of 3. Batch Normalization is again used to improve stability during training. Another Max Pooling layer with a pool size of 2 and a stride of 2 is applied. Following the convolutional and pooling layers, the output is flattened to a 1D vector to be fed into the fully connected layers. The fully connected layer with 32 neurons is applied to capture deeper representations of the extracted features. Finally, the model outputs a probability distribution over two classes using a softmax activation function, providing classification probabilities for NSTP and STP, respectively (Fig. 2).
The model was trained using the Adam optimizer from the Keras 3.5 library with a categorical cross-entropy loss function [13]. Training was conducted for 1,500 epochs with a batch size of 32 and a learning rate of 0.000003.
Test/validation
The model with the smallest validation error was chosen as the final model. This final model was then evaluated on the test set using the following metrics: accuracy, sensitivity, specificity, F1-score and area under the receiver operating curve (AUC).
Results
We sampled 89 HD treatments with SaO2 measurements from 22 in-center patients across four different clinics (Table 1). Continuous variables are reported as mean (± standard deviation) or median [Q1, Q3] and categorical variables are presented as proportions. In total, 4,075 consecutive 5-minute segments were adjudicated, 21.9% with STP and 78.1% with NSTP. The training dataset comprised 3,260 segments, while the test set contained 815 segments. Within the training set, NSTP and STP segments were distributed as 2,542 (78.0%) and 718 (22.0%), respectively. In the test set, these distributions were 642 (78.8%) for NSTP and 173 (21.2%) for STP (Table 2).
Our 1D-CNN model reached a 96.4% accuracy and an AUC of 0.995 (95% CI: 0.993 to 0.998) in the test set (Fig. 3). The respective confusion matrix is shown in Table 3. In the test data, the model reached 96.5% sensitivity, 96.4% specificity and 0.92 F1-score, indicating that it can work as a reliable method to detect STP.
For external validation, we randomly selected 20 recent HD treatments from 18 clinics. For this sampling, we did not perform the up-sampling of STPs to ensure a more accurate representation of their actual prevalence. Within this external validation dataset, the model achieved a sensitivity of 95.1%, a specificity of 96.7%, and an F1-score of 0.79 (Table 4).
Figure 4 shows the SaO2 measurements of two different HD treatments. Both sessions were about four hours long. In Fig. 4A, NSTP was classified, while for the session shown in Fig. 4B the model classified STP for many of the 5-minute segments (shown in orange).
To distinguish presumably benign from more concerning STP, we decided to further categorize STP in mild and severe STP. If a 5-minute segment was classified as STP and the minimum SaO2 in that segment was below 90% then we classified it as severe STP. If a 5-minute segment was classified as STP and all SaO2 measurements were equal or above 90% then we classified it as mild STP. Figure 5 shows the same HD treatment as Fig. 4B after separating mild and severe STP.
Discussion
Our primary finding demonstrates that a 1D-CNN can classify quasi-continuous intradialytic SaO2 recordings with high accuracy into STP and NSTP, respectively. The method can be scaled easily, enabling efficient classification of a large volume of SaO2 measurements. Given the prevalence and clinical significance of hypoxemia in HD patients, this result may hold important clinical implications.
While adequate SaO2 is fundamentally important for health and quality of life, there is a notable knowledge gap about SaO2 dynamics during HD. To address this issue, dialysis providers have implemented CLM devices in about 600 dialysis clinics across the US, many of which are connected to the cloud for seamless, quasi-continuous data gathering. Recently, we collected over six billion intradialytic SaO2 measurements from over 70,000 HD patients [14]. The analysis of such a vast data repository necessitates automated methods, which led us to develop an AI-based tool, a 1D-CNN. This model accurately classifies STP in SaO2 time series recorded during HD. We plan to deploy the model in the cloud, focusing on quantifying the SaO2 pattern at both the treatment and patient levels. This approach aims to enhance our understanding of intradialytic SaO2 dynamics and their clinical correlates.
SaO2 is a critical parameter in the management of patients undergoing HD, as it provides insights into the patient’s respiratory and circulatory status. In HD patients, maintaining adequate SaO2 is essential due to the unique physiological challenges posed by HD. Studies have shown that a significant proportion of hemodialysis patients experience episodes of hypoxia, with some reporting SaO2 levels dropping below 90% during treatment [1]. SaO2 serves as a prognostic indicator, with studies demonstrating that lower arterial oxygen saturation levels during HD are associated with poorer clinical outcomes, including higher rates of hospitalization and mortality [1]. In addition, low SaO2 during HD has been associated with intradialytic hypertension, possibly due to activation of the sympathetic nervous system activation, stimulation of the renin-angiotensin-aldosterone system (RAAS), and endothelin-1 release [6]. During the Covid pandemic, low SaO2 before the diagnosis of Covid-19 identified patients at high risk for hospitalization and death [3].
However, to date, the relationship between SaO2 STP and outcomes in HD patients is unknown. The reasons for this knowledge gap are multifaceted. First, it was only a few years ago that cloud-connected CLM devices were deployed, enabling the necessary data storage. Second, these newer CLM devices report SaO2 every 10 s, a significant improvement over the previous devices, which reported only once per minute. Nevertheless, it is conceivable that the ability to classify SaO2 STP could inform clinical decision-making or diagnostic and therapeutic interventions. For example, SaO2 STP are associated with SAS. This pattern typically reflects the intermittent nature of oxygen desaturation events that occur during apneic episodes. Galuzio et al. [4] highlight that oximetry data in HD patients can reveal intermittent oxygen desaturations that correlate with the apnea–hypopnea Index (AHI), a key metric used to assess the severity of sleep apnea. Furthermore, the relationship between SaO2 and SAS is underscored by findings from Jhamb et al. [15], who reported that lower mean oxygen saturation is associated with increased mortality risk in patients with advanced kidney disease, many of whom may also suffer from undiagnosed SAS. In addition, SAS in HD patients is associated with inflammation and oxidative stress [16]. In HD patients with SaO2 STP, specific clinical exploration and testing for SAS could be considered. If confirmed, e.g., through polysomnography, appropriate treatment could be initiated. This is relevant, as sleep-disordered breathing is frequent in dialysis patients [7]. The reported prevalence of sleep apnea is > 50% among dialysis patients [17,18,19]. SAS is a condition that is associated with lower health-related quality of life [20] and increased cardiovascular and mortality risk in kidney patients [15, 21].
Other methods have been developed to identify SaO2 STP. Galuzio et al. [4] devised a method that builds on dynamical systems theory and not on AI. The authors have developed a metric to quantify the occurrence of abnormal behaviors in the time series. This approach achieves an area under the ROC curve between 0.93 and 0.94 to identify SaO2 pattern indicative of SAS. It will be interesting to see the relative performance of the two approaches in a well-defined patient cohort with a clinical diagnosis of SAS.
When implemented at scale, the 1D-CNN will allow us to identify SaO2 STP and to investigate clinical correlates of the SaO2 STP, including demographic factors, laboratory data as well as treatment-related parameters. Additionally, one can examine the relationship between quantifiable measures of the SaO2 STP and key clinical outcomes, including hospitalization and mortality. This correlation seeks to provide deeper insights into the clinical significance of these patterns.
The quantification of STP at treatment level and patient level could also be used as feature inputs for other predictive models, e.g. hospitalization predictive model. With current cloud setting, this saw-tooth pattern recognition could be performed in real-time during the dialysis treatment and trigger a more specific diagnostic evaluation (e.g., polysomnography for definitive diagnosis of SAS) and therapeutic interventions.
Conclusions
In summary, we have developed an AI-based tool capable of detecting irregularities in SaO2 time series collected during hemodialysis. This tool will deepen our understanding of intradialytic SaO2 dynamics and support the implementation of advanced diagnostic and therapeutic interventions.
Data availability
As this was an internal quality improvement project, data will not be shared publicly.
Abbreviations
- SAS:
-
Sleep apnea syndrome
- SaO2 :
-
Arterial oxygen saturation
- HD:
-
Hemodialysis
- AI:
-
Artificial intelligence
- NSTP:
-
No saw-tooth pattern
- STP:
-
Saw-tooth pattern
- 1D-CNN:
-
One-dimensional convolutional neural networks
- PIH:
-
Prolonged intradialytic hypoxemia
- CHF:
-
Congestive heart failure
- COPD:
-
Chronic obstructive pulmonary disease
- CVD:
-
Cardiovascular disease
- ESKD:
-
End-stage kidney disease
- εopt :
-
Optimal recurrence threshold
- ECG:
-
Electrocardiograms
- EEG:
-
Electroencephalograms
- CLM:
-
Crit-line monitor
- FDA:
-
US Food and Drug Administration
- AWS:
-
Amazon web services
- AUC:
-
Area under the receiver operating characteristic curve
- CI:
-
Confidence interval
References
Meyring-Wosten A, et al. Intradialytic hypoxemia and clinical outcomes in patients on hemodialysis. Clin J Am Soc Nephrol. 2016;11(4):616–25.
Campos I, et al. Intradialytic hypoxemia in chronic hemodialysis patients. Blood Purif. 2016;41(1–3):177–87.
Preciado P, et al. Arterial oxygen saturation and hypoxemia in hemodialysis patients with COVID-19. Clin Kidney J. 2021;14(4):1222–8.
Galuzio PP, et al. Identification of arterial oxygen intermittency in oximetry data. Sci Rep. 2022;12(1):16023.
Zoccali C, Mallamaci F, Tripepi G. Nocturnal hypoxemia predicts incident cardiovascular complications in dialysis patients. J Am Soc Nephrol. 2002;13(3):729–33.
Meyring-Wosten A, et al. Intradialytic hypertension is associated with low intradialytic arterial oxygen saturation. Nephrol Dial Transpl. 2018;33(6):1040–5.
Unruh ML, et al. Sleep apnea in patients on conventional thrice-weekly hemodialysis: comparison with matched controls from the sleep heart health study. J Am Soc Nephrol. 2006;17(12):3503–9.
Pisano A, et al. Sleep apnoea syndrome prevalence in chronic kidney disease and end-stage kidney disease patients: a systematic review and meta-analysis. Clin Kidney J. 2024;17(1):sfad179.
Torino C, et al. Snoring amplifies the risk of heart failure and mortality in dialysis patients. Am J Nephrol. 2014;39(6):536–42.
Kiranyaz S, et al. 1-D Convolutional Neural Networks for Signal Processing Applications. 2019;8360–8364.
Apache Kafka. Available from: https://kafka.apache.org/
Keras. Available from: https://keras.io/.
Adam optimizer. Available from: https://keras.io/api/optimizers/adam/
Ban AN, et al. #861 intradialytic oxygen saturation in chronic hemodialysis patients—results from a large U.S. Population. Nephrol Dialys Transpl. 2024;39(Supplement_1).
Jhamb M, et al. Association of sleep apnea with mortality in patients with advanced kidney disease. Clin J Am Soc Nephrol. 2020;15(2):182–90.
Nikitidou O, et al. Sleep apnea syndrome, inflammation and oxidative stress in hemodialysis patients. Hemodial Int. 2018;22(2):209–16.
Stepanski E, et al. Sleep disorders in patients on continuous ambulatory peritoneal dialysis. J Am Soc Nephrol. 1995;6(2):192–7.
Kimmel PL, Miller G, Mendelson WB. Sleep apnea syndrome in chronic renal disease. Am J Med. 1989;86(3):308–14.
Wadhwa NK, Mendelson WB. A comparison of sleep-disordered respiration in ESRD patients receiving hemodialysis and peritoneal dialysis. Adv Perit Dial. 1992;8:195–8.
Gottlieb DJ, Punjabi NM. Diagnosis and management of obstructive sleep apnea: A review. JAMA. 2020;323(14):1389–400.
Puthenpura MM, et al. Chronic kidney disease and concomitant sleep apnea are associated with increased overall mortality: a meta-analysis. Int Urol Nephrol. 2020;52(12):2337–43.
Acknowledgements
We would like to thank the regulatory team of Renal Research Institute LLC for its support.
Funding
This research was funded out of the operational budget of the Renal Research Institute LLC. No external funding was received. H.Z. and P.K. are employees of Renal Research Institute LLC. A.N.B. is a contractor of Renal Research Institute LLC.
Author information
Authors and Affiliations
Contributions
H.Z. and P. K. developed the method and H.Z. and A.N.B. worked on the analysis. H.Z., P. K. and A.N.B. wrote the manuscript, H.Z. and A.N.B. prepared the figures and all authors reviewed the manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
The study was reviewed by the Western Institutional Review Board-Copernicus Group (WCG® IRB). The WCG® IRB determined that this research is exempt under 45 CFR § 46.104(d)(4) and that no patient consent was needed.
Competing interests
P. Kotanko and H. Zhang are employees and A. Nandorine Ban is a contractor of the Renal Research Institute, a wholly owned subsidiary of Fresenius Medical Care. P. Kotanko holds stock in Fresenius Medical Care. P. Kotanko and H. Zhang are inventors on patents in the field of dialysis. P. Kotanko is on the Editorial Board of Blood Purification, Kidney and Blood Pressure Research, and Frontiers in Nephrology.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.
About this article
Cite this article
Zhang, H., Nandorine Ban, A. & Kotanko, P. Intermittent hypoxemia during hemodialysis: AI-based identification of arterial oxygen saturation saw-tooth pattern. BMC Nephrol 26, 214 (2025). https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12882-025-04133-z
Received:
Accepted:
Published:
DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12882-025-04133-z