Pathological perspective reveals a novel hemodialyzer reaction: a case report

Abstract

Background

While the appearance of red clots in the dialyzer and the arterial and venous blood tubing lines is a common phenomenon in every hemodialysis unit, the occurrence of recurrent yellowish-white matter formation in the hemodialysis venous blood pot of a patient is rare.

Case presentation

We describe a male 69-year-old male with recurrent yellowish-white matter formation in the hemodialysis venous blood pot and red clots in the dialyzer. This was associated with a significant decrease in his red blood cells count. He had no history of thrombus no pro-thrombotic risk factors could be identified. Light microscopic examination of the deposits revealed the presence of large aggregates of neutrophils, large amounts of fibrin. The yellowish-white matter recurred at the next dialysis session. The occurrence of this episode was completely resolved by switching the dialysis filter and could not be avoided by increasing low molecular weight heparin dosage.

Conclusion

The yellowish-white matter and clotting within the dialyzer, as well as severe anemia, could be prevented by changing the type of dialyzer. Due to the rarity of this dialyzer reaction, it is important that awareness of this reaction by early identification be undertaken.

Peer Review reports

Introduction

The dialyzer is composed of many hollow fibers made from a biocompatible membrane across which solutes are cleared through diffusion and convection. In patients with uremia, after the blood is filtered through the hemodialyzer, some toxins can be removed, and the balance of water electrolytes and acid-base can be maintained, so that the function of the kidney is partially replaced, and only a very small amount of blood remains in the extracorporeal circulatory dialyzer and lines at the end of the dialysis treatment. The hemocompatibility of dialyzers for extracorporeal kidney replacement therapy is of importance to minimize harmful reactions between blood constituents and the membrane [1]. Nowadays, dialyzer reactions are hypersensitivity reactions to the membrane itself or the products used to sterilize the membrane. Dialyzer reactions have been characterized as type A or type B. Type A reactions occur early in the treatment, usually within the first 20 to 30 min. They typically occur during the first treatment but can occur after multiple treatments. Signs and symptoms may include pruritus, urticarial, laryngeal edema, bronchospasm, dyspnea, chest pain, vomiting, hypoxia, hypotension, and cardiac arrest. Management of a severe reaction includes stopping dialysis without returning blood from the extracorporeal circuit back into the patient. Type B reactions occur later in the treatment and are less severe. Symptoms may include chest and back pain, nausea, and vomiting [2, 3]. In patients who receive hemodialysis, most hypersensitivity reactions to components of the dialysis circuit are due to ethylene oxide or complement activating bio-incompatible membranes [4]. Dialysis is a life-sustaining procedure; therefore, prompt identification and management of the underlying cause of dialysis intolerance are crucial.

We found a novel type of reaction during hemodialysis, which the filter of the venous pot of the extracorporeal circulatory line was repeatedly clogged with a large amount of an unidentified substance and the hemodialyzer repeatedly clotted. Each dialysis was inadequate and extracorporeal blood was repeatedly lost. It was crucial to search for a cause and a solution to the problem of the production of this unidentified substance.

Case presentation

We present the case of a 69-year-old male weighing 62 kg, who was previously diagnosed with type 2 diabetes, diabetic nephropathy, chronic kidney disease stage 5, and uremia. The patient underwent hemodialysis three times weekly with low-molecular-weight heparin anticoagulation (4,000 U per session), utilizing an F16 dialyzer (WEGOBP; Hollow Fiber F16, polysulfone membrane) and an extracorporeal blood circuit for blood purification equipment (WEGOBP, consistent across all patients). Hemodialysis parameters included a blood flow rate of 230–250 mL/min and an average ultrafiltration rate of 400 mL/h. It is worth mentioning that this patient was using an F16 dialyzer for the first 2 weeks. There was no grade 3 coagulation in the dialyzer and no yellow-white deposits in the venous pot. Starting from the third week, each dialysis lasted 0.5–1 h and the hemodialysis machine frequently displayed a bubble alarm, accompanied by elevated venous pressure, and the dialysis was terminated prematurely. As a result, the hemodialyzer was observed to be grade 3 clotting (Fig. 1A) and a large amount of yellowish-white unidentified material was visible in the venous pot (Fig. 2B and C), which were deposited or adhered to the pot filters. The patient did not have any uncomfortable symptoms or abnormal signs, such as low blood pressure, rash, and chest tightness. Laboratory measurements are shown in Table 1. The patient’s hemoglobin decreased from 90 g/L at the beginning of dialysis to 53 g/L after 3 months of dialysis, which indicated severe anemia. The patient’s blood lipids were normal, coagulation indices were normal, the number of leukocytes and the number and ratio of neutrophils and eosinophils in the routine blood work were no obvious abnormalities. Blood biochemistry and clotting function were all within acceptable limits. However, no clinical risk factors could be identified. Thus, the pathologic investigation was performed to elicit the cause of yellowish-white matter in the venous pot.

Fig. 1
figure 1

Clotting was evaluated after hemodialysis with F16 filter (A) or with 17G filter (E) (black arrows)

Fig. 2
figure 2

Yellowish-white matter found adherent to the venous pot after rinse back with F16 dialyzer (black arrows, B and C) and matter disappeared with 17G (blue arrows, F)

Table 1 Relevant laboratory values

We fixed the unknown material in the venous pot using 4% formalin and sent it to the Pathology Department. The specimen was subjected to H&E staining (hematoxylin and eosin staining), which revealed abundant neutrophils and erythrocytes, large amounts of fibrin, and scattered individual lymphocyte (Fig. 3D and Fig. S2a). In addition, regarding analysis of blood immune cells, we also conducted immune-histochemical staining, as shown in Fig. S1. CD4 staining is negative. CD8 staining is negative. CD20 staining positive cells are extremely rare. CD68 positive cells are scattered in distribution. Based on the pathological findings, we considered that the patient’s blood contact with the membrane surface of the hemodialyzer might have produced a microinflammatory reaction. Then, after replacing the F16 dialyzer with a 17G dialyzer (NIPRO, SUREFLUX-17G, cellulose triacetate membrane), the patient’s hemodialysis dialyzer no longer coagulated (Fig. 1E) and the substance in the venous pot also completely disappeared (Fig. 2F). HE staining showed minimal amounts of neutrophils and fibrin (Fig. 3G, Fig. S2b). The patient’s hemoglobin level gradually returned to the normal range.

Fig. 3
figure 3

The light microscopic examination of the yellowish-white deposits in the venous pot with F16 (D) or 17G (G) showing large aggregates of neutrophils (black arrows) and erythrocytes (green arrows), large amounts of fibrin (blue arrows) (Hematoxylin-eosin stain, x400)

Discussion

Hemodialysis is one of the major life-sustaining treatments for patients with end-stage renal disease. Our patient had been receiving hemodialysis as an outpatient for 2 weeks without any complications. Only during subsequent hemodialysis, he developed repeated episodes of large amount of yellowish-white material in the venous pot within 0.5–1 h of starting the dialysis treatment. Repeated coagulation of the dialyzer seriously affects the quality of dialysis in patients. We were frankly at a loss to explain initially this clinic entity phenomenon.

The dialyzer reactions were clinically classified into type A and type B dialyzer reactions. Most type A reactions have been attributed to ethylene oxide, formaldehyde, and use of ACEIs with acrylonitrile dialyzers, with ethylene oxide accounting for most of these reactions. Ethylene oxide and formaldehyde cause a true, immunoglobulin E (IgE)-mediated anaphylaxis production [5]. Cuprophane and polysulfones/polyethersulfones appear to activate complement, which is thought to be the main cause of type B reactions. It was related to free hydroxyl groups on the membrane that activate the alternate complement pathway leading to neutrophil activation and pulmonary leukocyte sequestration. Our patient did not have any clinical symptoms, probably suggesting he had no IgE mediated hypersensitivity allergic response and activated complement pathway. Our patient only showed repeated coagulation of the dialyzer, and yellowish-white material repeatedly appeared in venous pot. This is a new type of dialyzer reaction, that which has not yet been reported yet. We judged this phenomenon by pathological methods as a reaction between the blood and the surface of the dialyzer membrane, which manifested itself as a large number of neutrophils and fibrin aggregation. Therefore, we only changed different types of dialysis membrane materials to solve this phenomenon, which we dialyzed him with a cellulose triacetate membrane based dialyzer and this adverse dialyzer reaction completely resolved. So, this patient should avoid exposure to polysulfone just as anyone allergic to peanuts should avoid any exposure to that antigen. However, the exact mechanism of the reaction between the patient’s blood and the dialysis membrane material is unknown and needs to be further studied. Neutrophil aggregation, a critical process in inflammatory responses and immune defense, is regulated by cytokines, chemokines, and intercellular interactions. The mechanism involved include: (1) chemokine- and cytokine-mediated recruitment, where IL-8 (CXCL8) binds to CXCR1/CXCR2 receptors to direct neutrophils to inflammatory sites [6]; TNF-α and IL-1β activate endothelial cells to upregulate adhesion molecules (e.g., ICAM-1, VCAM-1), thereby enhancing neutrophil adhesion and trans-endothelial migration [7]; Neutrophil-derived LTB4 (leukotriene B4) further amplifies aggregation via autocrine positive feedback [8]; (2) adhesion molecule-dependent aggregation, initiated by selectins (P-/E-selectin) mediating neutrophil rolling, followed by integrins (e.g., CD11b/CD18, Mac-1) binding endothelial ICAM-1 to stabilize adhesion and clustering [9, 10]. (3) NETs (neutrophil extracellular traps)-driven positive feedback [11]: Activated neutrophils release DNA protein complexes through NETs, capturing pathogens and activating more neutrophils. NETs components such as histone H3 (H3Cit) and elastase activate endothelial cells and platelets, inducing further secretion of IL-8 and TNF-α to promote neutrophil aggregation. As previously mentioned, the blood cell response is assumed to involve some cytokines and we don’t rule out all possible cytokine involvement in the blood circuit. If we observe yellow-white deposits during dialysis in other patient in the future, additional experiments such as cytokine quantification and in vitro modeling will be needed to conduct to further explore the mechanism underlying dialyzer response.

Adverse reactions with biocompatible polisulphone membranes are not frequent [12,13,14]. MD Arenas et al. described that the severe clinical syndrome of broncoespasm during hemodialysis with several biocompatible polysulphone membranes made by different manufacturers and with a variety of sterilization methods. On following day he was dialyzed on an celulose triacetate dialyzer and the hemodialysis treatment was uneventful [12]. W H Boer ea al. presented two cases of dialyser reactions, both in patients using a polysulfone. Patient 1 suffered from recurrent attacks of acute dyspnoea, hypoxia and hypotension that occurred early in dialysis sessions, whereas patient 2 presented with unexplained episodes of severe hypotension and vomiting in the initial phases of dialysis [15]. Among 159 patients in our hemodialysis center, only 1 patient was observed to have yellow white substance in venous pot during hemodialysis with a polysulfone membrane. The patient was transitioned to the celulose triacetate membrane dialyzer, which he tolerated well and continued to utilize through the followed hemodialysis course without complication. These cases remind us that “biocompatible” membranes are not free from dialyzer reactions. It also demonstrates a wide range of clinical dialyzer reaction presentations and the complex nature involving a sensitivity reaction to hemodialyzer.

We believe that this is a new type of dialyzer reaction that deserves attention and that pathology plays an important diagnostic role. For clinical cases in which the dialyzer still coagulates or venous pressure is elevated after anticoagulation, rather than simply increasing the anticoagulant and thus the risk of bleeding, this type of novel dialyzer reaction should be taken into consideration. In the future, we believe that improvements in the biocompatibility of dialysis membranes will reduced biological responses elicited by blood-membrane interactions.

Data availability

Data sharing is not applicable to this article as no datasets were generated or analysed during the current study.

Abbreviations

ACEIs:

Angiotensin-converting enzyme inhibitors

ICAM-1:

Intercellular adhesion molecule-1

VCAM-1:

Vascular cell adhesion molecule-1

References

  1. Wagner S, et al. Hemocompatibility of polysulfone Hemodialyzers - Exploratory studies on impact of treatment modality and dialyzer characteristics. Kidney360. 2020;1:25–35.

    Article  PubMed  PubMed Central  Google Scholar 

  2. Greenberg KI, Choi MJ. Hemodialysis emergencies: Core curriculum 2021. Am J Kidney Dis. 2021;77:796–809.

    Article  PubMed  Google Scholar 

  3. Saha M, Allon M. Diagnosis, treatment, and prevention of Hemodialysis emergencies. Clin J Am Soc Nephrol. 2017;12:357–69.

    Article  CAS  PubMed  Google Scholar 

  4. Sayeed K, Murdakes C, Spec A, Gashti C. Anaphylactic shock at the beginning of Hemodialysis. Semin Dial. 2016;29:81–4.

    Article  PubMed  Google Scholar 

  5. Kopacz A, Ludwig C, Tarbox M. Atypical cutaneous and musculoskeletal manifestation of SARS-CoV-2: ‘COVID-19 toes’ and spasticity in a 48-year-old woman. BMJ Case Rep. 2021;14.

  6. Carnicelli D, et al. The antibiotic polymyxin B impairs the interactions between Shiga Toxins and Human Neutrophils. J Immunol. 2016;196:1177–85.

    Article  CAS  PubMed  Google Scholar 

  7. Ye J, et al. Zebrafish as a model for investigating Klebsiella pneumoniae-driven lung injury and therapeutic targets. Exp Lung Res. 2025;51:11–22.

    Article  CAS  PubMed  Google Scholar 

  8. Kawasaki Y, et al. The leukotriene B4 receptor antagonist ONO-4057 inhibits mesangioproliferative changes in anti-Thy-1 nephritis. Nephrol Dial Transpl. 2005;20:2697–703.

    Article  CAS  Google Scholar 

  9. Kruger P, et al. Neutrophils: Between host defence, immune modulation, and tissue injury. PLoS Pathog. 2015;11:e1004651.

    Article  PubMed  PubMed Central  Google Scholar 

  10. Walzog B, et al. A role for beta(2) integrins (CD11/CD18) in the regulation of cytokine gene expression of polymorphonuclear neutrophils during the inflammatory response. FASEB J. 1999;13:1855–65.

    Article  CAS  PubMed  Google Scholar 

  11. Wigerblad G, Kaplan MJ. Neutrophil extracellular traps in systemic autoimmune and autoinflammatory diseases. Nat Rev Immunol. 2023;23:274–88.

    Article  CAS  PubMed  Google Scholar 

  12. Arenas MD, Gil MT, Carreton MA, Moledous A, Albiach B. [Adverse reactions to polysulphone membrane dialyzers durind hemodialysis]. Nefrologia. 2007;27:638–42.

    CAS  PubMed  Google Scholar 

  13. Watnick S, Stooksbury M, Winter R, Riscoe M, Cohen DM. White thrombus formation in blood tubing lines in a chronic Hemodialysis unit. Clin J Am Soc Nephrol. 2008;3:382–6.

    Article  PubMed  PubMed Central  Google Scholar 

  14. Sathe KP, et al. Recurrent white thrombi formation in Hemodialysis tubing: A case report. BMC Nephrol. 2015;16:3.

    Article  PubMed  PubMed Central  Google Scholar 

  15. Boer WH, Liem Y, de Beus E, Abrahams AC. Acute reactions to polysulfone/polyethersulfone dialysers: literature review and management. Neth J Med. 2017;75:4–13.

    CAS  PubMed  Google Scholar 

Download references

Acknowledgements

None.

Funding

This work was supported by grants from the Pudong New Area Science and Technology Development Fund Public Institution Livelihood Research Special Project (PKJ2024-Y07 to Weijuan Lou), Shanghai Sixth People’s Hospital Basic Research General Cultivation Project (ynms202305 to Jianyong Yin) and Shanghai Sixth People’s Hospital Hospital Management Research Center-Hospital Management Health Consortium Research Special Project (lylht202313 to Yongping Guo).

Author information

Authors and Affiliations

Authors

Contributions

WJL and YPG conceptualized, designed and NSW supervised the study. YCX, TFC, JYY and YJT contributed to the data collection, such as information collection, scanning, and data input and WJL and JF Ganalyzed the data. XlC and YL provided essential images. WJL wrote the manuscript. JHD, YFY supported data collection. All authors reviewed the manuscript.

Corresponding author

Correspondence to Weijuan Lou.

Ethics declarations

Ethics approval and consent to participate

Written informed consent to participate obtained from patient. A copy of the written informed consent is available for review by the Editor of this journal.

Consent for publication

The patient provided written informed consent for the publication of this case report.

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Electronic supplementary material

Below is the link to the electronic supplementary material.

Supplementary Material 1

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Lou, W., Xi, Y., Liu, Y. et al. Pathological perspective reveals a novel hemodialyzer reaction: a case report. BMC Nephrol 26, 246 (2025). https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12882-025-04180-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12882-025-04180-6

Keywords